Figure 6: Whole slide image scanners with different scan coverage. An H and E stained slide with several large portions of tissue is shown in which one fragment at the top of the slide extends beyond the coverslip. (Left) The black and white macro image correctly shows all tissue pieces including the portion extending beyond the coverslip; the corresponding whole slide image accurately scanned all of the tissue (Hamamatsu HT2). (Right) The macro image is shown with a green box identifying the tissue region to be scanned, including the tissue extending beyond the coverslip; the corresponding scanned whole slide image is missing part of the tissue near the upper edge of the slide (Aperio AT2)

Figure 6: Whole slide image scanners with different scan coverage. An H and E stained slide with several large portions of tissue is shown in which one fragment at the top of the slide extends beyond the coverslip. (Left) The black and white macro image correctly shows all tissue pieces including the portion extending beyond the coverslip; the corresponding whole slide image accurately scanned all of the tissue (Hamamatsu HT2). (Right) The macro image is shown with a green box identifying the tissue region to be scanned, including the tissue extending beyond the coverslip; the corresponding scanned whole slide image is missing part of the tissue near the upper edge of the slide (Aperio AT2)